Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hosp...
7.5CVSS
7.3AI Score
0.002EPSS
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hosp...
9.8CVSS
9.2AI Score
0.002EPSS
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recomme...
9.8CVSS
9.5AI Score
0.002EPSS
Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network...
9.8CVSS
9.3AI Score
0.002EPSS